GENERAL INFORMATION |
|
PRESSURE SENSOR SPECIAL FEATURES |
|
MEASURING RANGES |
|
PRESSURE SENSOR PERFORMANCE |
Accuracy | ≤ ± 0.04 %FS, ≤ ± 0.055 %FS
| Accuracy Details | ■ Reference Accuracy of Calibrated Span (includes the eff ects of terminal-based linearity, hysteresis, and repeatability) ■ ±0.055% of Span ■ ±0.04% of Span (with /HAC option)
| Non-linearity | N/A
| Non-linearity Details | N/A
| Long Term Stability | ≤ ± 0.1 %FS
| Long Term Stability Details | Stability (All normal operating condition): ±0.1% of URL per 10 years
| TC Sensitivity | ± (0.083% Span + 0.02% URL) ± (0.083% Span + 0.014% URL)
| TC Zero Point | ± (0.083% Span + 0.02% URL) ± (0.083% Span + 0.014% URL)
| Adjustable | YES
| Adjustability Details | ■ Primary Variable: 100:1 ■ Parameter configuration by the external zero adjustment screw and push button (Integral indicator code E) offers easy and quick setup for parameters.
| Resolution | External zero is continuously adjustable with 0.01% incremental resolution of span.
| Sensitivity | N/A
| Response Time | < 150 ms, < 90 ms
| Response Time Details | Response Time (All capsules): ■ 90 ms ■ 150 ms for H capsule with Wetted Parts Material code H, M, T, A, D, B and W.
|
---|
|
ELECTRICAL SPECIFICATIONS |
Output Signal | 1 … 5 VDC HART®, 4 … 20 mA (BRAIN protocol), 4 … 20 mA HART®, FOUNDATION™ Fieldbus, PROFIBUS® PA
| Output Signal Details | ■ 4-20 mA (BRAIN protocol) ■ 4-20 mA (HART 5/HART 7 protocol) ■ FOUNDATION Fieldbus protocol, refer to GS 01C31T02-01EN)
| Supply Voltage | N/A
| Load Resistance | N/A
| Power Consumption | N/A
| Short-circuit Protection | N/A
| Reverse Polarity Protection | N/A
|
---|
|
ENVIROMENTAL CONDITIONS |
Media Temperature [°C] [Min] | -40
| Media Temperature [°C] [Max] | +120
| Ambient Temperature [°C] [Min] | -40
| Ambient Temperature [°C] [Max] | +85
| Storage Temperature [°C] [Min] | N/A
| Storage Temperature [°C] [Max] | N/A
| Compensated Temperature [°C] [Min] | N/A
| Compensated Temperature [°C] [Max] | N/A
| Ingress Protection | IP66, IP67, NEMA 4X
| Vibration Resistance | N/A
| Shock Resistance | N/A
|
---|
|
GENERAL MECHANICAL SPECIFICATIONS |
Mounting | Mounting Bracket, Threaded – Internal Diaphragm
| Process Connection Male/Female | Female
| Process Connection | 1/2" NPT, 1/4" NPT, Adapters, Cover Flange, Rc 1/2", RC 1/4", Valve Manifold
| Process Connection Details | N/A
| With Damper (Puls-snubber) | NO
| Process Connection Material | Hastelloy® C-276 (2.4819), Monel, Stainless steel 316 (ASTM CF-8M), Stainless steel 316L (ASTM CF-3M), Super duplex Stainless steel
| Diaphragm / Sensor Material | Hastelloy® C-276 (2.4819), Monel, Stainless steel (316L), Stainless steel F316L (ASTM A182), Tantalum
| Filling Fluid | Fluorinated® oil, Silicon oil
| Seal | PTFE, Teflon-coated 316L SST
| Hygienic Version | NO
| Electrical Connection | Field housing with cable gland 1/2" NPT, Field housing with cable gland G 1/2", Field housing with cable gland M20x1.5
| Electrical Connection Details | Refer to the datasheet
| Display | YES
| Display Details | 5-digit numerical LCD display, 6-digit unit display and bargraph. The indicator is configurable to display one or up to three of the following variables periodically.;
| Weight | [Installation code 7, 8 and 9] 2.8 kg(6.2 lb) for wetted parts material code S and L, without integral indicator, mounting bracket, and
|
---|
|
TESTS / APPROVALS |
Scope Approvals | ATEX / Europe, CE / European Community, CSA / Canada, EAC / Eurasian Economic Community, FM / USA, INMETRO / Brazil, KCs (KOSHA) / South Korea, NEPSI / China, Regulatory Compliance Mark (RCM) / Australia
| Ship Approvals | ABS / USA, BV / Global, DNV / Germany, LR / The UK
| Explosion Protection | Dust Explosion Protection Through Housing "t", Flameproof Enclosure "d", Intrinsic Safety "i", Non-incendive "n"
| EX Zone | N/A
| Other Approvals | N/A
| SIL Qualification | SIL 2, SIL 3
|
---|
|
INDUSTRIES / APPLICATIONS |
Industry | Chemical, Machine building, Mechanical engineering & Automation, Mining, minerals and metals, Oil and gas, Petrochemical, Water and wastewater
| Application | Calibration & testing labs, Liquid, steam and gas pressure measurement, Oil field equipment, Oil platforms, Pressure monitoring in the hazardous areas, Process technology
|
---|
|